Case Narrative — SDG CLE-2026-SDG-0142
========================================

Laboratory:   Clearline Environmental Laboratory
              1234 Lab Way, Blackshear, GA 31516
              NELAC #NELAC-2024-CLE-001

Client:       RJ Lee Group
              Pittsburgh, PA / Pasco, WA

Project:      2026-Q2 Groundwater + Hanford IH Air
SDG number:   CLE-2026-SDG-0142
Date issued:  2026-06-16

Samples received (2026-06-15):
  GW-0009   MW-01 Groundwater (Pittsburgh)
  GW-0010   MW-02 Groundwater (Pittsburgh) -- See note 1
  MB-0003   Method Blank
  AF-0005   Air filter (Hanford 200E-01)
  AF-0006   Air filter (Hanford 200E-02)

Analyses performed:
  EPA 6020B  ICP-MS metals panel (23 analytes), water samples
  NIOSH 7300 Metals on MCE air filters, air samples
  EPA 7470A  Mercury (Cold Vapor AA), water samples

Sample receipt:
  All samples received within holding times, properly preserved,
  receipt temperature 4.2 C (within 0-6 C range).

QC summary:
  Method Blank:        All analytes < MDL
  LCS recovery:        98.2% (limits 80-120%)
  MS / MSD recovery:   96.4% / 95.0% (limits 75-125%)
  RPD:                 1.4% (limit <20%)
  All QC within acceptance criteria.

Note 1 -- GW-0010 Zinc reanalysis:
  Initial analysis (injection #16, 10:24) reported counts
  exceeding the high calibration standard (62,485,200 cps).
  Sample was automatically queued for dilution reanalysis at
  x10 (injection #18, 11:24), bracketed by CCV-2 to confirm
  calibration stability. The dilution-corrected result
  (12,400 ug/L, flag D) supersedes the original (>5000 ug/L,
  flag E) per CLP SOM02.4 section 10.2. Both runs are
  preserved in the audit trail and reported in Stage 2+
  packages.

Anomalies, deviations, or corrective actions:
  None other than note 1 above.

Approved by:
  M. Vasquez (Analyst)         2026-06-16 14:30
  K. Patel (QA Manager)         2026-06-16 14:32
  Clearline Laboratory Director Signature: _______________